IntroductionNon-invasive monitoring of oxygenation has become a standard procedure. Of the two methods available, namely transcutaneous partial pressure of oxygen (tcpO2) monitoring and pulse oximetry, the former is currently at risk of being replaced by the latter. Show
This is despite the fact that both methods provide substantially different information about oxygenation and are particularly helpful if used in combination. This article reviews the principles of both techniques and discusses methodological and practical issues related to the use of these devices in infants. Transcutaneous partial pressure of oxygen (tcpO2) monitoringPrinciple of operation Electrodes are heated to improve oxygen diffusion and to arterialize the capillary blood. Oxygen is reduced at the cathode, generating an electric current proportional to the O2 concentration in the capillary bed underneath the sensor. Sensors require a 10-15 min warm-up period after application and have to be calibrated every 4-8 hours. FIG 1. Structure of a combined tcpO2/pCO2 transcutaneous electrode. Factors influencing measurements Thus, transcutaneous (tcpO2) and arterial (pO2(aB)) represent two different measurements, and at times changes in tcpO2 are more important than the actual values measured. In hemodynamically unstable children, tcpO2 will reflect changes in the circulatory status and thereby give an early warning of the onset of circulatory compromise. Sensor temperature This is particularly important in preterm neonates, in whom high pO2(aB) levels must be reliably detected to minimize the risk of retinopathy of prematurity. Probe placement In infants with a closed arterial duct, the electrode should ideally be placed on the lower back or abdomen, or on a thigh. Peripheral perfusion If an underreading of pO2(aB) occurs, it is advisable to check the patient for these conditions [2]. Skin thickness Studies in children and adults did in fact show that the ratio between tcpO2 and pO2(aB) is extremely constant in these patients (independent of age and pO2(aB)), it merely is 20 % lower than in neonates, i.e. approximately 0.8 [3]. Detection of hypoxemia and hyperoxemia In this regard, the data available suggest that approximately 15 % of both hypoxemic and hyperoxemic instances are missed by tcpO2 monitors, whereas their specificity, particularly with regard to hypoxemia, is somewhat higher [5]. Pulse oximetry (sO2(p))Principle of operation The ratio of the absorbance of red and infrared light sent through a tissue correlates with the proportion of oxygenated to deoxygenated hemoglobin in the tissue. Conventional pulse oximeters determine the arterial component within this absorbance by identifying the peaks and troughs in the absorbance over time, thereby obtaining a "pulse-added" absorbance that is independent of the absorbance characteristics of the non-pulsating parts of the tissue. These pulse-added light absorbances are then associated algorithmically with empirically determined arterial oxygen saturation (sO2(aB)) values [1]. FIG 2. Conventional pulse oximetry : Light passes through the layers of the measuring site. When the light absorption at the trough (which should include all the constant absorbers) is subtracted from the light absorption at the peak then, in theory, the resultants are the absorption characteristics due to added volume of arterial blood only. Next-generation instruments use additional and/or different techniques. For example, one recently developed technology scans through all red-to-infrared ratios (and corresponding sO2(p) values) found in the tissue, determines the intensity of these and chooses the right-most peak of these intensities, which will correspond to the absorbance by the arterial blood in the tissue. It also uses frequency analysis, time domain analysis and adaptive filtering to establish a 'noise reference' in the detected physiological signal [6], thereby improving the ability to separate between signal and noise (see below). Other next-generation instruments use differential signal amplification to achieve this goal. Factors influencing measurements However, it is probably because of this apparent ease of use that potentially erroneous measurements on a pulse oximeter are more at risk of being overlooked than those occurring with a tcpO2monitor. A thorough understanding of the factors potentially affecting the precision of a pulse oximeter is therefore particularly important [1]. Probe placement Light bypassing the tissue can cause both falsely high and falsely low values. The sensor site must be checked every 6-8 hours. It was recently shown that 42 % of nurses in a neonatal intermediate-care nursery exceeded a pressure on the skin of 50 mmHg/6.7 kPa during fixation of a pulse oximeter sensor [7]. Such high pressures may result in a reduced signal-to-noise ratio and may thus severely impair the precision of the sO2(p) measurements [7]. Highly flexible sensors provide better skin contact and thus better signal-to-noise ratio. Peripheral perfusion Because next-generation oximeters rely less on pulse detection, they continue to operate even at lower blood pressure levels [9]. Response times However, all pulse oximeters currently available average their values over 2-15 sec to level out any erroneous measurement which may occasionally occur even under optimal conditions. This averaging, however, has unwanted consequences:
Motion artifact The pulsatile (=arterial) component contributes less than 1 % to the total absorbance measured by the pulse oximeter. Hence, at least conventional pulse oximetry is very sensitive to sudden changes in background signal, e.g. due to body movements. Next-generation instruments use various techniques to identify and read through periods with low signal-to-noise ratios as there are during motion, resulting in dramatic (> 90 %) decreases in false alarm rates [10,11]. However, some of these improvements are achieved at the expense of not identifying true desaturation [12], which is unacceptable. With conventional oximeters, identification of motion artifact is best done by displaying the light plethysmographic waveforms. Whenever these are distorted, sO2(p) readings become unreliable. Alternatively, the pulse rate can be compared with the heart rate from an ECG monitor, which should be identical [13]. For next-generation instruments, which are less reliant on a clean peak-and-trough detection and thus an undisturbed pulse waveform, there is currently no independently validated method to identify periods of poor measurement conditions and thus potentially unreliable sO2(p) readings. An interesting approach in this regard is the signal quality indicator, recently shown reliably to exclude erroneous measurements if above a certain threshold [14]. Other hemoglobins and pigments Fetal hemoglobin (HbF) and bilirubin do not affect pulse oximeters but may lead to a bias on sO2(aB) by CO-oximeters. In patients with dark skin, sO2(p) values may be falsely high, particularly during hypoxemia [1]. Algorithms These may vary between brands and even between different software versions from the same manufacturer. Also, some instruments subtract a priori the typical levels of COHb, MetHb, etc. in healthy non-smoking adults from their measurements and will thus displaysO2(p) values that are some 2-3 % lower than those displayed by other instruments. This approach, i.e. to display the so-called "fractional" sO2(p) instead of the usual "functional" sO2(p), has largely been abandoned in recent years, probably because it resulted in an unacceptably poor ability of instruments using this approach to detect hyperoxemia [15]. FIG. 3. The Oxygen Dissociation Curve illustrates the relationship between sO2(aB) and pO2(aB). Various conditions can shift the curve left or right as described in the boxes. Detection of hypoxemia and hyperoxemia Due to the shape of the O2 dissociation curve, however, they are less well suited for detecting hyperoxemia. The upper alarm limits that have to be chosen on individual instrument brands to avoid hyperoxemia reliably range from 88 to 95 % [1,15]. An upper alarm limit of 95 % was recently confirmed for three next-generation instruments [16]. The reliability in detecting sO2(p) values >80 mmHg/10.7 kPa via non-invasive monitoring can likely be increased if both sO2(p) and tcpO2 are monitored. tcpO2
or pulse oximetry? For example, at high sO2(aB) levels, large changes in pO2(aB) may occur with very small changes in sO2(aB). Also, at a constant pO2(aB) of, for example, 45 mmHg/6 kPa, sO2(aB) may be either 80 % or 88 %, depending on whether arterial pH is 7.25 to 7.40. The actual relationship between sO2(aB) and pO2(aB) and/or between tcpO2 and sO2(p) should, therefore, be repeatedly determined, particularly in critically ill patients. In addition, both tcpO2 monitors and pulse oximeters have their specific shortfalls: the tcpO2 monitor is relatively difficult to use, has frequent periods where there is no reliable measurement (during warm-up and calibration periods) and reacts relatively slowly to sudden changes in oxygenation, whereas at least conventional pulse oximeters fail to operate reliably whenever the patient moves, are relatively prone to erroneous measurements resulting from improper probe placement and may be unreliable under conditions of hyperoxemia and possibly also severe hypoxemia. For these reasons, both devices should ideally be used in combination, particularly whenever a reliable detection of hyperoxemia is paramount, as it is in extremely low-birthweight infants receiving additional inspired oxygen. In these patients, it may be advisable to reduce the FIO2 whenever one of the two instruments alarms. Disclaimer May contain information that is not supported by performance and intended use claims of Radiometer's products. See also Legal info. When assessing a newborn for early signs of jaundice which technique does the nurse use?Can be checked by pressing one finger on a baby's forehead or nose (called "blanching" the skin). If the skin is jaundiced, it will appear yellow when the finger is removed. Can be followed in some babies by pressing over their bony prominences of their chest, hips, and knees to check if the jaundice is worsening.
When performing an assessment on a neonate which assessment finding is most suggestive of hypothermia quizlet?When performing an assessment on a neonate, which assessment finding is most suggestive of hypothermia? 1. Bradycardia. Hypothermic neonates become bradycardic proportional to the degree of core temperature.
Which assessment of the newborn should be reported quizlet?What symptom assessed in the newborn shortly after delivery should be reported? Sternal retractions are evidence that the newborn is in respiratory distress and should be reported immediately.
What is it called when an infant during the birth process experiences a restriction of oxygen lasting a few minutes that produces cognitive deficits?Birth asphyxia, the condition in which a baby is deprived of oxygen at birth, is a leading cause of infant brain damage. Brain damage is a serious health concern, leading to lifelong consequences for a child. Oxygen deprivation at birth can cause intellectual deficits and learning disabilities.
|